1 July 2004 Fiber-based Raman spectroscopy of glucose
Author Affiliations +
Abstract
Design and functionality of a reagent-free Raman-based fiber optic sensor are described in the context of glucose monitoring. Theoretical calculations of sensor performance are shown using raytrace software. An optimized sensor design with an outer diameter of 1.5 mm was carried out. First experiments in aqueous glucose solutions showed a root mean square error of prediction of 10.5 mg/dl in an independent validation.
© (2004) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Daniel Rohleder, Wolfgang Kiefer, Michael Schoemaker, Wolfgang H. Petrich, "Fiber-based Raman spectroscopy of glucose", Proc. SPIE 5321, Biomedical Vibrational Spectroscopy and Biohazard Detection Technologies, (1 July 2004); doi: 10.1117/12.528304; https://doi.org/10.1117/12.528304