16 June 2004 Optical biosensors based on four different surface plasmon resonance modes: sensitivity comparison
Author Affiliations +
Abstract
Current surface plasmon resonance (SPR) modes based on the attenuated total reflection (ATR) method can broadly be categorized as: conventional SPR, long-range SPR (LRSPR), coupled plasmon-waveguide resonance (CPWR), and waveguide-coupled SPR (WCSPR). Although the features of optical biosensors are dependent upon their particular SPR mode, a common requirement for all biosensors utilized for biomolecular interaction analysis (BIA) is a high degree of sensitivity. The current paper presents a theoretical analysis and comparison of the sensitivity and resolution of these four types of SPR biosensors when employed in three of the most prevalent detection methods, namely angular interrogation, wavelength interrogation, and intensity measurement. This study develops a detailed understanding of the influences of various biosensor design parameters in order to enhance the sensitivity and detection limit capabilities of such devices.
© (2004) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Fan-Ching Chien, Fan-Ching Chien, Shean-Jen Chen, Shean-Jen Chen, } "Optical biosensors based on four different surface plasmon resonance modes: sensitivity comparison", Proc. SPIE 5327, Plasmonics in Biology and Medicine, (16 June 2004); doi: 10.1117/12.530434; https://doi.org/10.1117/12.530434