5 May 2004 Tracked ultrasound for laparoscopic surgery
Philip Bao M.D., John R. Warmath, Benjamin Poulose M.D., Robert L. Galloway Jr., Alan J. Herline M.D.
Author Affiliations +
Abstract
We calibrate a tracked laparoscopic ultrasound probe for application in image-guided surgery and 3-D volume reconstruction. With a plane-mapping technique, the spatial relationship between the ultrasound beam emitted from the tip of the probe to the local coordinate system of the probe was determined by mapping it with an optically tracked pointer. A cross-wire calibration technique was also performed for comparison. The accuracy and precision of the calibrated probe was evaluated by measuring its ability to localize targets in a water bath. Target registration error depended upon probe position, varying from an average 0.88 mm for the fixed probes to 6.09 mm for a moving probe. This error can be reduced to 4.54 mm by accounting for target localization error which is the error determining the position of the probe itself. These results validate the plane-mapping calibration technique for this type of ultrasound probe, and better probe tracking is expected to reduce the overall registration error.
© (2004) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Philip Bao M.D., John R. Warmath, Benjamin Poulose M.D., Robert L. Galloway Jr., and Alan J. Herline M.D. "Tracked ultrasound for laparoscopic surgery", Proc. SPIE 5367, Medical Imaging 2004: Visualization, Image-Guided Procedures, and Display, (5 May 2004); https://doi.org/10.1117/12.535429
Lens.org Logo
CITATIONS
Cited by 7 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Ultrasonography

Calibration

Laparoscopy

Image registration

Imaging systems

Surgery

Cameras

Back to Top