7 July 2004 An EIT reconstruction algorithm: comparison of one-step and iterative versions
Author Affiliations +
Proceedings Volume 5505, Optical Methods, Sensors, Image Processing, and Visualization in Medicine; (2004) https://doi.org/10.1117/12.577929
Event: Optical Methods, Sensors, Image Processing, and Visualization in Medicine, 2003, Gdansk, Poland
Abstract
A reconstruction algorithm based on applying of a current pattern using electrodes is considered in the paper. A resulting voltage distribution between electrodes is measured. To improve an accuracy of the reconstructed image two FEM models are used -- a dense one in the forward problem and coarse one in inverse problem. A one-step version of the algorithm is examined and the results are compared with these obtained from iterative version of the algorithm. The reconstruction algorithm based on Levenberg-Marquardt approach is utilized. It has been found that the errors of images depend on object properties and can vary significantly in one-step version of reconstruction algorithm.
© (2004) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Agnieszka Janczulewicz, Agnieszka Janczulewicz, Jerzy Wtorek, Jerzy Wtorek, } "An EIT reconstruction algorithm: comparison of one-step and iterative versions", Proc. SPIE 5505, Optical Methods, Sensors, Image Processing, and Visualization in Medicine, (7 July 2004); doi: 10.1117/12.577929; https://doi.org/10.1117/12.577929
PROCEEDINGS
7 PAGES


SHARE