10 February 2005 Analysis of alignment error in asphere testing using a corrector
Author Affiliations +
Abstract
The characteristics and effects of alignment error on test results are analysed in this paper. The mathematical model are established, the least square method is used to determine the misalignment. Procedures are presented that allow for separation of misalignment from misfigure in asphere testing. The misalignment-induced aberrations by corrector and the support of the mirror can be calculated and subtracted from segment interferogram so that only true misfigure remains. The accuracy of the system can reach up to 0.028λ(λ=632.8nm).
© (2005) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Donglin Xue, Ligong Zheng, and Xuejun Zhang "Analysis of alignment error in asphere testing using a corrector", Proc. SPIE 5638, Optical Design and Testing II, (10 February 2005); https://doi.org/10.1117/12.574991