Paper
17 January 2005 Effect of laser intensity changing on accuracy of size measurement by laser scanning
Author Affiliations +
Abstract
In the system of size measurement by laser scanning, the laser intensity is necessarily kept stability. But in practice, output intensity of laser is continuously attenuated as time passed, so that precision accuracy of detecting system becomes poor. The paper analyzes the effect of laser intensity changing on accuracy of size measurement by laser scanning, provides with the auto-measured circuit and the measuring methods of the output laser intensity, and the method of electrical compensation for intensity change. The method has been applied to the instrument of diameter measurement by laser scanning on line successfully, which is effective and feasible.
© (2005) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Tao Jiang, Zhenglin Yu, Guohua Cao, and Hongji Xu "Effect of laser intensity changing on accuracy of size measurement by laser scanning", Proc. SPIE 5644, Optoelectronic Devices and Integration, (17 January 2005); https://doi.org/10.1117/12.577452