13 April 2005 Mirror image free Fourier domain optical coherence tomography
Author Affiliations +
Abstract
A novel swept source based Fourier domain optical coherence tomography (FDOCT) system using an electro-optic phase modulator was demonstrated. The imaging range was doubled by cancellation of the mirror image. The elimination of low frequency noises due to DC and autocorrelation terms increases the sensitivity by 20 dB.
© (2005) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Jun Zhang, John Stuart Nelson, Zhongping Chen, "Mirror image free Fourier domain optical coherence tomography", Proc. SPIE 5690, Coherence Domain Optical Methods and Optical Coherence Tomography in Biomedicine IX, (13 April 2005); doi: 10.1117/12.592827; https://doi.org/10.1117/12.592827