20 April 2005 Characterization of scatter radiation in cone beam CT mammography
Author Affiliations +
Abstract
Cone beam CT mammography (CBCTM) is an emerging breast imaging technology and is currently under intensive investigation [1-3]. One of the major challenges in CBCTM is to understand the characteristics of scatter radiation and to find ways to reduce or correct its degrading effects. Since the breast shape, geometry and image formation process are significantly different from conventional mammography, all system components and parameters such as target/filter combination, kVp range, source to image distance, detector design etc. should be examined and optimized. In optimizing CBCTM systems, it is important to have knowledge of how different imaging parameters affect the recorded scatter within the image. In this study, a GEANT4 based Monte Carlo simulation package (GATE) was used to investigate the scatter magnitude and its’ distribution in CBCTM. The influences of different air gaps, kVp settings, breast sizes and breast composition on the scatter primary ratio (SPR) and scatter profiles were examined. In general, the scatter to primary ratio (SPR) is strongly dependent on the breast size and air gap, and is only moderately dependent on the kVp setting and breast composition. These results may be used for optimization of CBCTM systems, as well as for developing scatter correction methods.
© (2005) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Bob Liu, Stephen J. Glick, Corinne Groiselle, "Characterization of scatter radiation in cone beam CT mammography", Proc. SPIE 5745, Medical Imaging 2005: Physics of Medical Imaging, (20 April 2005); doi: 10.1117/12.595396; https://doi.org/10.1117/12.595396
PROCEEDINGS
10 PAGES


SHARE
Back to Top