20 April 2005 Determining MTF of digital detector system with Monte Carlo simulation
Eun Seon Jeong, Hyung Won Lee, Sang Hee Nam
Author Affiliations +
Abstract
We have designed a detector based on a-Se(amorphous Selenium) and done simulation the detector with Monte Carlo method. We will apply the cascaded linear system theory to determine the MTF for whole detector system. For direct comparison with experiment, we have simulated 139um pixel pitch and used simulated X-ray tube spectrum.
© (2005) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Eun Seon Jeong, Hyung Won Lee, and Sang Hee Nam "Determining MTF of digital detector system with Monte Carlo simulation", Proc. SPIE 5745, Medical Imaging 2005: Physics of Medical Imaging, (20 April 2005); https://doi.org/10.1117/12.594996