12 April 2005 Two-dimensional capacitive micromachined ultrasonic transducer (CMUT) arrays for a miniature integrated volumetric ultrasonic imaging system
Xuefeng Zhuang, Ira O. Wygant, David T. Yeh, Amin Nikoozadeh, Omer Oralkan, Arif S. Ergun, Ching-Hsiang Cheng, Yongli Huang, Goksen G. Yaralioglu, Butrus T. Khuri-Yakub
Author Affiliations +
Abstract
We have designed, fabricated, and characterized two-dimensional 16x16-element capacitive micromachined ultrasonic transducer (CMUT) arrays. The CMUT array elements have a 250-μm pitch, and when tested in immersion, have a 5 MHz center frequency and 99% fractional bandwidth. The fabrication process is based on standard silicon micromachining techniques and therefore has the advantages of high yield, low cost, and ease of integration. The transducers have a Si3N4 membrane and are fabricated on a 400-μm thick silicon substrate. A low parasitic capacitance through-wafer via connects each CMUT element to a flip-chip bond pad on the back side of the wafer. Each through wafer via is 20 μm in diameter and 400 μm deep. The interconnects form metal-insulator-semiconductor (MIS) junctions with the surrounding high-resistivity silicon substrate to establish isolation and to reduce parasitic capacitance. Each through-wafer via has less than 0.06 pF of parasitic capacitance. We have investigated a Au-In flip-chip bonding process to connect the 2D CMUT array to a custom integrated circuit (IC) with transmit and receive electronics. To develop this process, we fabricated fanout structures on silicon, and flip-chip bonded these test dies to a flat surface coated with gold. The average series resistance per bump is about 3 Ohms, and 100% yield is obtained for a total of 30 bumps.
© (2005) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Xuefeng Zhuang, Ira O. Wygant, David T. Yeh, Amin Nikoozadeh, Omer Oralkan, Arif S. Ergun, Ching-Hsiang Cheng, Yongli Huang, Goksen G. Yaralioglu, and Butrus T. Khuri-Yakub "Two-dimensional capacitive micromachined ultrasonic transducer (CMUT) arrays for a miniature integrated volumetric ultrasonic imaging system", Proc. SPIE 5750, Medical Imaging 2005: Ultrasonic Imaging and Signal Processing, (12 April 2005); https://doi.org/10.1117/12.594702
Lens.org Logo
CITATIONS
Cited by 15 scholarly publications and 3 patents.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Capacitance

Silicon

Transducers

Semiconducting wafers

Resistance

Etching

Ultrasonics

Back to Top