Abstract
A wavemeter based on Michelson interferometer accurately measure static wavelength of a tunable laser. Its operation principle is formulated in details. Double longitudinal-mode He-Ne laser with frequency stabilization is used as the reference optical source of the wavemeter. Voice-coil motor using PID means can realize to move in uniform motion. Phase-locked loop circuit including NE564 and 74LS193 is used to enhance resolution of the wavemeter. Data processing is carried out by the counter unit including two 8254 programmable timer, a MCU, a LCD. The test shows that its measurement accuracy is 1×10-6 and is higher than those of other wavemeters such as Fizeau interference and Fabry-Perot wavemeter.
© (2005) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Liqiang Wang, Liqiang Wang, Wenjie Ren, Wenjie Ren, } "Michelson interferometer for laser wavelength", Proc. SPIE 6020, Optoelectronic Materials and Devices for Optical Communications, 60202M (2 December 2005); doi: 10.1117/12.636227; https://doi.org/10.1117/12.636227