Abstract
Repeated micro-CT scanning of a number of iliac crest biopsies enabled us to quantitate the variation in CT image gray-scale and spatial geometry due to variables such as specimen orientation, projection magnification, voxel size and slight differences in x-ray photon energy in each of the different scans. Using the micro-CT scanner on beamline X2B at the Brookhaven National Laboratory's National Synchrotron Light Source, we rescanned several iliac crest bone biopsy specimens, and a test phantom made of calcium hydroxyapatite, at repeated scanning sessions and evaluated the reproducibility of the spatial geometry and gray-scale haracteristics of the specimens. This scanner consists of a Bragg diffiaction source of monochromatic x-rays, a computer controlled high precision specimen rotation and translation stage assembly, and a fluorescent crystal and CCD array system for imaging the specimen at each of the angles of view around its axis of rotation during the scanning sequence. The 3-D micro-CT images consisted of up to 1024x24002, 4 μm, cubic voxels, each with 16-bit gray-scale. We also reconstructed the images at 16,32 and 48 μm voxel resolution. Partial volume effects at the surface of the bone were diminished by 'eroding' the surface voxels in the 4 μm images, but significantly changed the outcome at greater voxel size. Reproducibility of the mineral content of bone, at mean bone opacity value, was ± 28.8 mg/cm3, i.e., 2.56%, in a 4 μm cubic voxel at the 95% confidence level.
© (2006) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Steven M. Jorgensen, Diane R. Eaker, Andrew J. Vercnocke, Erik L. Ritman, "Reproducibility of 3D micro-CT gray-scale and structural dimension data in longitudinal studies", Proc. SPIE 6143, Medical Imaging 2006: Physiology, Function, and Structure from Medical Images, 61433H (13 March 2006); doi: 10.1117/12.654436; https://doi.org/10.1117/12.654436
PROCEEDINGS
9 PAGES


SHARE
Back to Top