19 May 2006 Study on precise measurement of high reflectivity by cavity ring-down spectroscopy
Author Affiliations +
Abstract
Using cavity ring-down spectroscopy technique, the precise measurement device for the high reflectivity at 1.315μm has been built. Tested reflectivity of high-reflecting mirror is as high as 99.979%. The precision of measurement is at the 10-5 level. For further investigation, some factors such as the instability of laser source, the cavity length, the diaphragm aperture and detector position, the mode match, the reflectivity of cavity mirror, which may affect the precision of measurement are studied. Our study will provide a theoretical and experimental reference for further improvement of the measuring accuracy.
© (2006) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Gang Ren, Gang Ren, Bang-wei Cai, Bang-wei Cai, Bin Zhang, Bin Zhang, Sheng-ming Xiong, Sheng-ming Xiong, Wei Huang, Wei Huang, Li-feng Gao, Li-feng Gao, } "Study on precise measurement of high reflectivity by cavity ring-down spectroscopy", Proc. SPIE 6150, 2nd International Symposium on Advanced Optical Manufacturing and Testing Technologies: Optical Test and Measurement Technology and Equipment, 61500O (19 May 2006); doi: 10.1117/12.678091; https://doi.org/10.1117/12.678091
PROCEEDINGS
6 PAGES


SHARE