19 May 2006 InGaAs trap detector used as a near-infrared transfer standard for detector calibrations
Author Affiliations +
Abstract
This paper presents results of investigations of the radiant power spectral responsivity within the near-infrared spectral range. Large-area windowless InGaAs photodiodes of 10mm×10mm are studied in the spectrum ranged from 900 nm to 1600 nm. The InGaAs trap detector used as a transfer standard for power measurements has been characterized and calibrated, obtaining an uncertainty of less than 0.1% at several wavelengths.
© (2006) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Ji Wang, Ji Wang, Shuang Li, Shuang Li, Lei Zhang, Lei Zhang, Feng Chen, Feng Chen, Xiaobing Zheng, Xiaobing Zheng, } "InGaAs trap detector used as a near-infrared transfer standard for detector calibrations", Proc. SPIE 6150, 2nd International Symposium on Advanced Optical Manufacturing and Testing Technologies: Optical Test and Measurement Technology and Equipment, 61501D (19 May 2006); doi: 10.1117/12.676739; https://doi.org/10.1117/12.676739