12 June 1986 Image Quality In Selenium-Based Digital Radiography
Author Affiliations +
Proceedings Volume 0626, Application of Optical Instrumentation in Medicine XIV and Picture Archiving and Communication Systems; (1986) https://doi.org/10.1117/12.975390
Event: Application of Optical Instrumentation in Medicine XIV and Picture Archiving and Communication Systems (PACS IV) for Medical Applications, 1986, Newport Beach, CA, United States
Abstract
Capacitive sensing of x-ray generated charge images on amorphous selenium layers represents an attractive technique for the electronic recording of projection radiographs. In order to evaluate the practical limits of image quality with such a selenium detector we have set up a laboratory system that allows the recording of small size (70 x 70 mm2) radiographic images. We have measured MTF and noise power spectra and have calculated from these data noise equivalent quanta (NEQ) and detective quantum efficiency (DQE) as a function of dose and spatial frequency. On the basis of these quantities the imaging performance of the system is discussed.
© (1986) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
U. Schiebel, W. Hillen, and T. Zaengel "Image Quality In Selenium-Based Digital Radiography", Proc. SPIE 0626, Application of Optical Instrumentation in Medicine XIV and Picture Archiving and Communication Systems, (12 June 1986); https://doi.org/10.1117/12.975390