13 February 2007 Speed-of-sound imaging in a photoacoustic imager
Author Affiliations +
Abstract
The effects of an inappropriately chosen speed-of-sound in photoacoustic imaging reconstructions are to cause blurring of images and impairment of contrast. Here we outline a new methodology to measure the speed-of-sound in a photoacoustic imager with little or no additional cost and without the need to perform extra measurements. The method uses a strong absorber of light which is placed in the path of the light illuminating the sample. This acts as a source of ultrasound whose interaction with the sample can be measured at the far-end of the sample using the same ultrasound detector used for photoacoustics. This yields time-of-arrival measurements of the ultrasound transient at the multi-element detector. Such measurements are made at various angles around the sample in a computerized tomography approach. Reconstruction of the time-of-arrival or speed-of-sound tomogram of the object can be made by implementing a fan-beam projection reconstruction algorithm. We present the concept and validate the method on a speed-of-sound phantom.
© (2007) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Srirang Manohar, Srirang Manohar, René G. H. Willemink, René G. H. Willemink, Ton G. van Leeuwen, Ton G. van Leeuwen, } "Speed-of-sound imaging in a photoacoustic imager", Proc. SPIE 6437, Photons Plus Ultrasound: Imaging and Sensing 2007: The Eighth Conference on Biomedical Thermoacoustics, Optoacoustics, and Acousto-optics, 64370R (13 February 2007); doi: 10.1117/12.700078; https://doi.org/10.1117/12.700078