You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
16 March 2007Dual-energy technique for digital flat-panel detectors without x-ray tube voltage switching
Dual-energy imaging increases the possibility of pulmonary nodule detection by reducing the bone structure
noise. The major problem of the dual-energy acquisition process with digital flat-panel detectors is the interval
of time between low-energy (LE) exposure and high-energy (HE) exposure. Due to misregistration between LE
and HE images, motion artifacts pollute the subtracted image. This paper presents a new acquisition approach
for dual-energy imaging developed in order to reduce this inter-exposure time. The idea is to keep the tube
voltage constant and to just switch a filter in front of the imaged object and thus to modulate the outgoing x-ray
spectrum.
The first part of this study presents how to optimize system parameters for the new acquisition protocol:
source voltage, dynamic filtration before the patient, exposure time for LE and HE acquisition. The tube load
is kept constant to focus the optimization study on the dose and the exposure time. A noise quality factor
(NQF) and a spectral quality factor (SQF) are used as criteria for optimization. The new approach system is
then compared to the state-of-the-art system with voltage switching between low and high energy. A filtering
algorithm of dual energy acquisitions enabling a significant noise reduction is presented. Performance between its
combination with the new acquisition protocol and the reference one are compared. For a limited noise quality
factor, three times faster acquisition time is obtained using the new system. Noise reduction techniques improve
the image SNR by 61% in the new system and only 32% in the reference system, without taking into account
the impact of better registration on the dual-energy image quality.
The alert did not successfully save. Please try again later.
Christopher S. Coello, Jean-Marc Dinten, Marc Arques, Pierre Rohr, Christophe Odet, "Dual-energy technique for digital flat-panel detectors without x-ray tube voltage switching," Proc. SPIE 6510, Medical Imaging 2007: Physics of Medical Imaging, 651005 (16 March 2007); https://doi.org/10.1117/12.707355