16 March 2007 Configuration of AEC kVp dependence for digital radiography systems
Author Affiliations +
Abstract
As digital radiography (DR) systems are being increasingly adopted for clinical applications, automatic exposure control (AEC) has remained a critically important component. Unlike traditional screen/film systems, however, there does not currently exist a widely accepted AEC calibration criterion for DR systems. This is due mainly to the signal response characteristics and wide dynamic range of a DR detector, which are inherently different from those of a screen/film system. Consequently, the AEC cutoff dose and its dependence on the kVp selection (i.e., kVp compensation) should be calibrated differently for DR systems. In this paper, we have investigated three possible schemes to set up the AEC compensation based on a constant response of the detector in terms of the signal, receptor dose, or signal-to-noise-ratio (SNR) respectively. The results for each of the setup schemes were evaluated on four different DR detectors (Gd2O2S, CsI(Tl), a-Se, or BaFBrI as x-ray absorption material) based on the measured signal and noise response of the detector under the ISO beam conditions (ISO 6236-1). The results showed that all three setup schemes produced similar results for clinical beams above 70 kVp. Significant differences were observed only at lower kVp (≤60) beams. In addition, schemes of constant signal and constant SNR produced similar results with the only exception for the a-Se detector at low kVp (≤60) beam. These results indicate that the choice of the kVp schemes would be important only for low kVp exams.
© (2007) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Weidong Huang, Weidong Huang, Richard Van Metter, Richard Van Metter, Chang-Ying Joseph Yang, Chang-Ying Joseph Yang, John Yorkston, John Yorkston, } "Configuration of AEC kVp dependence for digital radiography systems", Proc. SPIE 6510, Medical Imaging 2007: Physics of Medical Imaging, 651017 (16 March 2007); doi: 10.1117/12.714377; https://doi.org/10.1117/12.714377
PROCEEDINGS
10 PAGES


SHARE
RELATED CONTENT

Noise Sources In Digital Fluorography
Proceedings of SPIE (November 03 1981)
Linear system models for lag in flat dynamic x ray...
Proceedings of SPIE (April 19 2005)
Megalopinakophobia: its symptoms and cures
Proceedings of SPIE (June 27 2001)

Back to Top