20 March 2007 Visual image quality metrics for optimization of breast tomosynthesis acquisition technique
Author Affiliations +
Abstract
Breast tomosynthesis is currently an investigational imaging technique requiring optimization of its many combinations of data acquisition and image reconstruction parameters for optimum clinical use. In this study, the effects of several acquisition parameters on the visual conspicuity of diagnostic features were evaluated for three breast specimens using a visual discrimination model (VDM). Acquisition parameters included total exposure, number of views, full resolution and binning modes, and lag correction. The diagnostic features considered in these specimens were mass margins, microcalcifications, and mass spicules. Metrics of feature contrast were computed for each image by defining two regions containing the selected feature (Signal) and surrounding background (Noise), and then computing the difference in VDM channel metrics between Signal and Noise regions in units of just-noticeable differences (JNDs). Scans with 25 views and exposure levels comparable to a standard two-view mammography exam produced higher levels of feature contrast. The effects of binning and lag correction on feature contrast were found to be generally small and isolated, consistent with our visual assessments of the images. Binning produced a slight loss of spatial resolution which could be compensated in the reconstruction filter. These results suggest that good image quality can be achieved with the faster and therefore more clinically practical 25-view scans with binning, which can be performed in as little as 12.5 seconds. Further work will investigate other specimens as well as alternate figures of merit in order to help determine optimal acquisition and reconstruction parameters for clinical trials.
© (2007) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Jeffrey P. Johnson, Joseph Lo, Thomas Mertelmeier, John S. Nafziger, Pontus Timberg, and Ehsan Samei "Visual image quality metrics for optimization of breast tomosynthesis acquisition technique", Proc. SPIE 6515, Medical Imaging 2007: Image Perception, Observer Performance, and Technology Assessment, 65150P (20 March 2007); https://doi.org/10.1117/12.712343
Lens.org Logo
CITATIONS
Cited by 2 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Visualization

Interference (communication)

Breast

Data acquisition

Image resolution

Visual process modeling

Diagnostics

Back to Top