You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
12 April 2007Retinal images in the human eye with implanted intraocular lens
A typical proceeding in cataract is based on the removal of opaque crystalline lens and inserting in its place the artificial
intraocular lens (IOL). The quality of retinal image after such procedure depends, among others, on the parameters of the
IOL, so the design of the implanted lens is of great importance. An appropriate choice of the IOL material, especially in
relation to its biocompatibility, is often considered. However the parameter, which is often omitted during the IOL design is
its chromatic aberration. In particular lack of its adequacy to the chromatic aberration of a crystalline lens may cause
problems.
In order to fit better chromatic aberration of the eye with implanted IOL to that of the healthy eye we propose a hybrid - refractive-diffractive IOL. It can be designed in such way that the total longitudinal chromatic aberration of an eye with
implanted IOL equals the total longitudinal chromatic aberration of a healthy eye.
In this study we compare the retinal image quality calculated numerically on the basis of the well known Liou-Brennan
eye model with typical IOL implanted with that obtained if the IOL is done as hybrid (refractive-diffractive) design.
The alert did not successfully save. Please try again later.
Marek Zając, Damian Siedlecki, Jerzy Nowak, "Retinal images in the human eye with implanted intraocular lens," Proc. SPIE 6609, 15th Czech-Polish-Slovak Conference on Wave and Quantum Aspects of Contemporary Optics, 66091D (12 April 2007); https://doi.org/10.1117/12.739679