Paper
28 November 2007 Particle size measurement system of the aerosol time-of-flight mass spectrometer optimization study
Wen-wu Zhao, Xue-jun Gu, Li-qing Hao, Ming-qiang Huang, Hai-yang Zheng, Li Fang, Wei-jun Zhang
Author Affiliations +
Abstract
In this paper, Aerosol time-of-flight mass spectrometer optical particle size measurement system optimization and influencing factors are studied. Optimized the distance between nozzle and the skimmer, using a shorter wavelength light source measurement, Aerosol particle size measurement from the previous limit of 260 nm to 120 nm. Study found that the particles scattered light intensity is connected with light intensity and wavelength of the light source. particle size measurements using shorter laser wavelengths or greater energy, which can measure smaller size diameter within certain range. combining particle beam focusing aerodynamic lens system device, the particle size of air the time-resolved measurement of 256 / min, Measurement of time-resolved mass spectrometry to 170 / min, MS hit rate of 64.5%.
© (2007) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Wen-wu Zhao, Xue-jun Gu, Li-qing Hao, Ming-qiang Huang, Hai-yang Zheng, Li Fang, and Wei-jun Zhang "Particle size measurement system of the aerosol time-of-flight mass spectrometer optimization study", Proc. SPIE 6834, Optical Design and Testing III, 683422 (28 November 2007); https://doi.org/10.1117/12.755311
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Atmospheric particles

Particles

Aerosols

Light scattering

Spectroscopy

Laser scattering

Aerodynamics

Back to Top