19 February 2008 Correction for source decay in bioluminescence tomography
Author Affiliations +
Abstract
Time variation of the bioluminescence source can cause artifacts in the tomographic images. We showed that the a priori knowledge of the light kinetics could be used to eliminate these artifacts. We performed two-dimensional simulations. We considered a 40-mm-diameter circular region with an inclusion of 6-mm-diameter located 10-mm away from the center. The measurement data was simulated using a finite element based forward solver. We modeled the non-contact measurements such that four-wavelength data was collected from four 90-degree-apart views. The results showed that the ratio of the total imaging time to the half-life of the exponentially decaying bioluminescent source was the deciding factor in the reconstruction of the source. It was also demonstrated that a priori knowledge of the source kinetics was required to perform tomographic bioluminescence imaging of short half-life bioluminescent sources and the use of spatial a priori information alone was not adequate.
© (2008) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.