Abstract
Shorter pulses, in theory, should be favorable in nonlinear microscopy and yield stronger signals. However, shorter pulses are much more prone to chromatic dispersion when passing through the microscope objective, which significantly broadens its pulse duration and cancels the expected signal gain. In this paper, multiphoton intrapulse interference phase scan (MIIPS) was used to compensate chromatic dispersion introduced by the 1.45 NA objective. The results show that with MIIPS compensation, the increased signal is realized. We also find that third and higher order dispersion compensation, which cannot be corrected by prism pairs, is responsible for an additional factor of 4.7 signal gain.
© (2008) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Peng Xi, Peng Xi, Lindsay R. Weisel, Lindsay R. Weisel, Yair Andegeko, Yair Andegeko, Vadim V. Lovozoy, Vadim V. Lovozoy, Marcos Dantus, Marcos Dantus, } "Two-photon laser scanning microscopy with ultrabroad bandwidth 110 nm FWHM femtosecond pulses", Proc. SPIE 6860, Multiphoton Microscopy in the Biomedical Sciences VIII, 68601U (15 February 2008); doi: 10.1117/12.763698; https://doi.org/10.1117/12.763698