11 March 2008 Automated motion correction based on target tracking for dynamic nuclear medicine studies
Author Affiliations +
Abstract
Nuclear medicine dynamic studies of kidneys, bladder and stomach are important diagnostic tools. Accurate generation of time-activity curves from regions of interest (ROIs) requires that the patient remains motionless for the duration of the study. This is not always possible since some dynamic studies may last from several minutes to one hour. Several motion correction solutions have been explored. Motion correction using external point sources is inconvenient and not accurate especially when motion results from breathing, organ motion or feeding rather than from body motion alone. Centroid-based motion correction assumes that activity distribution is only inside the single organ (without background) and uniform, but this approach is impractical in most clinical studies. In this paper, we present a novel technique of motion correction that first tracks the organ of interest in a dynamic series then aligns the organ. The implementation algorithm for target tracking-based motion correction consists of image preprocessing, target detection, target positioning, motion estimation and prediction, tracking (new search region generation) and target alignment. The targeted organ is tracked from the first frame to the last one in the dynamic series to generate a moving trajectory of the organ. Motion correction is implemented by aligning the organ ROIs in the image series to the location of the organ in the first image. The proposed method of motion correction has been applied to several dynamic nuclear medicine studies including radionuclide cystography, dynamic renal scintigraphy, diuretic renography and gastric emptying scintigraphy.
© (2008) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Xinhua Cao, Tracy Tetrault, Fred Fahey, Ted Treves, "Automated motion correction based on target tracking for dynamic nuclear medicine studies", Proc. SPIE 6914, Medical Imaging 2008: Image Processing, 69142B (11 March 2008); doi: 10.1117/12.770440; https://doi.org/10.1117/12.770440
PROCEEDINGS
9 PAGES


SHARE
RELATED CONTENT

The algorithm of small and weak linear target detection
Proceedings of SPIE (September 08 2011)
Screening closing scenarios for tactical targets
Proceedings of SPIE (July 01 2002)
Efficient approach to physics-based FOPEN-SAR ATD/R
Proceedings of SPIE (June 10 1996)
Coherent aspect-dependent SAR image formation
Proceedings of SPIE (June 09 1994)
Image sequence recognition
Proceedings of SPIE (September 16 1994)

Back to Top