Paper
3 April 2008 A segmentation method for stentgrafts in the abdominal aorta from ECG-gated CTA data
Almar Klein, W. KlaasJan Renema, Luuk J. Oostveen, Leo J. Schultze Kool M.D., Cornelis H. Slump
Author Affiliations +
Abstract
Endovascular aortic replacement (EVAR) is an established technique, which uses stentgrafts to treat aortic aneurysms in patients at risk of aneurysm rupture. The long-term durability of a stentgraft is affected by the stresses and hemodynamic forces applied to it, and may be reflected by the movements of the stentgraft itself during the cardiac cycle. A conventional CT scan (which results in a 3D volume) is not able to visualize these movements. However, applying ECG-gating does provide insight in the motion of the stentgraft caused by hemodynamic forces at different phases of the cardiac cycle. The amount of data obtained is a factor of ten larger compared to conventional CT, but the radiation dose is kept similar for patient safety. This causes the data to be noisy, and streak artifacts are more common. Algorithms for automatic stentgraft detection must be able to cope with this. Segmentation of the stentgraft is performed by examining slices perpendicular to the centreline. Regions with high CT-values exist at the locations where the metallic frame penetrates the slice. These regions are well suited for detection and sub-pixel localization. Spurious points can be removed by means of a clustering algorithm, leaving only points on the contour of the stent. We compare the performance of several different point detection methods and clustering algorithms. The position of the stent's centreline is calculated by fitting a circle through these points. The proposed method can detect several stentgraft types, and is robust against noise and streak artifacts.
© (2008) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Almar Klein, W. KlaasJan Renema, Luuk J. Oostveen, Leo J. Schultze Kool M.D., and Cornelis H. Slump "A segmentation method for stentgrafts in the abdominal aorta from ECG-gated CTA data", Proc. SPIE 6916, Medical Imaging 2008: Physiology, Function, and Structure from Medical Images, 69160R (3 April 2008); https://doi.org/10.1117/12.770064
Lens.org Logo
CITATIONS
Cited by 11 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Image segmentation

Detection and tracking algorithms

Computed tomography

Image processing algorithms and systems

Metals

Databases

Hemodynamics

Back to Top