6 March 2008 Optimizing breast-tomosynthesis acquisition parameters with scanning model observers
Author Affiliations +
Abstract
In breast tomosynthesis (BT), multiple x-ray projections obtained over a limited angular span are reconstructed to produce a three-dimensional (3D) volume. This 3D imagery can lead to reduced structural masking effects compared to conventional mammography. Accordingly, there has been considerable interest in optimizing acquisition and reconstruction parameters associated with BT. In this work, we evaluate the use of a scanning model observer and localization ROC (LROC) methodology for performing a task-based optimization of the angular span and number of projection angles for a simulated BT system. The observer was applied to extracted slices of 3D volumes reconstructed with filtered backprojection. Both "background-known-exactly" (BKE) and "quasi-BKE" (QBKE) tasks were conducted. The latter task attempts to account for limited observer training by preserving structural noise in the detection task. Reduced noise in the form of fewer projections was important with the BKE task, although wider angular spans were also advantageous. Higher sampling densities may improve performance for the more-realistic QBKE tasks.
© (2008) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
H. C. Gifford, C. S. Didier, Mini Das, and S. J. Glick "Optimizing breast-tomosynthesis acquisition parameters with scanning model observers", Proc. SPIE 6917, Medical Imaging 2008: Image Perception, Observer Performance, and Technology Assessment, 69170S (6 March 2008); https://doi.org/10.1117/12.771018
Lens.org Logo
CITATIONS
Cited by 14 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Breast

Tumors

3D acquisition

Mammography

Tumor growth modeling

3D modeling

Reconstruction algorithms

Back to Top