Paper
2 May 2008 New optical analyzer for 13C-breath test
Hermann Harde, Matthias Dressler, Günther Helmrich, Marcus Wolff, Hinrich Groninga
Author Affiliations +
Abstract
Medical breath tests are well established diagnostic tools, predominantly for gastroenterological inspections, but also for many other examinations. Since the composition and concentration of exhaled volatile gases reflect the physical condition of a patient, a breath analysis allows one to recognize an infectious disease in an organ or even to identify a tumor. One of the most prominent breath tests is the 13C-urea-breath test, applied to ascertain the presence of the bacterium helicobacter pylori in the stomach wall as an indication of a gastric ulcer. In this contribution we present a new optical analyzer that is based on photoacoustic spectroscopy and uses a DFB diode laser at 2.744 μm. The concentration ratio of the CO2 isotopologues is determined by measuring the absorption on a 13CO2 line in comparison to a 12CO2 line. In the specially selected spectral range the lines have similar strengths, although the concentrations differ by a factor of 90. Therefore, the signals are well comparable. Due to an excellent signal-noise-ratio isotope variations of less than 1% can be resolved as required for the breath test.
© (2008) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Hermann Harde, Matthias Dressler, Günther Helmrich, Marcus Wolff, and Hinrich Groninga "New optical analyzer for 13C-breath test", Proc. SPIE 6991, Biophotonics: Photonic Solutions for Better Health Care, 69910T (2 May 2008); https://doi.org/10.1117/12.781271
Lens.org Logo
CITATIONS
Cited by 3 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Modulation

Absorption

Photoacoustic spectroscopy

Carbon dioxide

Semiconductor lasers

Carbon monoxide

Molecules

RELATED CONTENT

Towards laser-based open-path detection of hydrogen sulfide
Proceedings of SPIE (December 23 2016)
Photoacoustic CO<sub>2</sub> detection at 2.7 &#956;m
Proceedings of SPIE (April 25 2006)
Optoacoustic 13C-breath test analyzer
Proceedings of SPIE (February 23 2010)
Photoacoustic HF sensor
Proceedings of SPIE (April 28 2008)

Back to Top