Paper
25 March 1976 Thermal Characteristics Of Rotating Anode X-Ray Tubes
Thomas Koller
Author Affiliations +
Proceedings Volume 0070, Application of Optical Instrumentation in Medicine IV; (1976) https://doi.org/10.1117/12.954570
Event: Application of Optical Instrumentation in Medicine, IV, 1975, Atlanta, United States
Abstract
The criteria determining thermal loading of x-ray tubes is reviewed. Major parameters of maximum rated heat storage capacity and maximum continuous dissipation are defined as related to the thermal characteristics of the x-ray tube target and its radiographic loadability. The resulting anode thermal characteristic chart is proposed and provides for a 25% target storage capacity value on which to base radiographic ratings. For procedures in which the target heat storage exceeds 25% of the maximum rated storage capacity, a reduction to 0.8 of the maximum radiographic load rating is proposed. A target design capable of withstanding greater thermal loads and results of qualification tests are reviewed.
© (1976) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Thomas Koller "Thermal Characteristics Of Rotating Anode X-Ray Tubes", Proc. SPIE 0070, Application of Optical Instrumentation in Medicine IV, (25 March 1976); https://doi.org/10.1117/12.954570
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
X-rays

Molybdenum

Tungsten

Medicine

Surgery

Diagnostics

Rhenium

RELATED CONTENT

Equipment Installation Acceptance
Proceedings of SPIE (December 23 1976)
The Loadability Of Rotating Anode X-Ray Tubes
Proceedings of SPIE (March 25 1976)
Large Area Contrast Prediction In Screen-Film Systems
Proceedings of SPIE (August 18 1980)
Quality Assurance: A Fundamental Whose Time Has Come
Proceedings of SPIE (March 25 1976)

Back to Top