A precision centrifuge is used to test and calibrate accelerometer model parameters. Its dynamic unbalance may cause the
perturbation of the centrifuge to deteriorate the test and calibration accuracy of an accelerometer. By analyzing the
causes of dynamic unbalance, the influences on precision centrifuge from static unbalance and couple unbalance are
developed. It is considered measuring and balancing of static unbalance is a key to resolving a dynamic unbalance
problem of precision centrifuge with a disk in structure. Measuring means and calculating formulas of static unbalance
amount are given, and balancing principle and method are provided. The correctness and effectiveness of this method are
confirmed by experiments on a device under tuning, thereby the accurate and high-effective measuring and balancing
method of dynamic unbalance of this precision centrifuge was provided.
Yafei Yang andXin Huo
"Measuring and balancing dynamic unbalance of precision centrifuge", Proc. SPIE 7133, Fifth International Symposium on Instrumentation Science and Technology, 71330T (12 January 2009); https://doi.org/10.1117/12.808588
ACCESS THE FULL ARTICLE
INSTITUTIONAL Select your institution to access the SPIE Digital Library.
PERSONAL Sign in with your SPIE account to access your personal subscriptions or to use specific features such as save to my library, sign up for alerts, save searches, etc.
The alert did not successfully save. Please try again later.
Yafei Yang, Xin Huo, "Measuring and balancing dynamic unbalance of precision centrifuge," Proc. SPIE 7133, Fifth International Symposium on Instrumentation Science and Technology, 71330T (12 January 2009); https://doi.org/10.1117/12.808588