Translator Disclaimer
12 March 2009 Characterization of a novel photon counting detector for clinical CT: count rate, energy resolution, and noise performance
Author Affiliations +
Proceedings Volume 7258, Medical Imaging 2009: Physics of Medical Imaging; 725824 (2009) https://doi.org/10.1117/12.813915
Event: SPIE Medical Imaging, 2009, Lake Buena Vista (Orlando Area), Florida, United States
Abstract
We report on a characterization study of a multi-row direct-conversion x-ray detector used to generate the first photon counting clinical x-ray computed tomography (CT) patent images. In order to provide the photon counting detector with adequate performance for low-dose CT applications, we have designed and fabricated a fast application specific integrated circuit (ASIC) for data readout from the pixellated CdTe detectors that comprise the photon counting detector. The cadmium telluride (CdTe) detector has 512 pixels with a 1 mm pitch and is vertically integrated with the ASIC readout so it can be tiled in two dimensions similar to those that are tiled in an arc found in 32-row multi-slice CT systems. We have measured several important detector parameters including the maximum output count rate, energy resolution, and noise performance. Additionally the relationship between the output and input rate has been found to fit a non-paralyzable detector model with a dead time of 160 nsec. A maximum output rate of 6 × 106 counts per second per pixel has been obtained with a low output x-ray tube for CT operated between 0.01 mA and 6 mA at 140 keV and different source-to-detector distances. All detector noise counts are less that 20 keV which is sufficiently low for clinical CT. The energy resolution measured with the 60 keV photons from a 241Am source is ~12%. In conclusion, our results demonstrate the potential for the application of the CdTe based photon counting detector to clinical CT systems. Our future plans include further performance improvement by incorporating drift structures to each detector pixel.
© (2009) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
William C. Barber, Einar Nygard, Jan S. Iwanczyk, Mengxi Zhang, Eric C. Frey, Benjamin M. W. Tsui, Jan C. Wessel, Nail Malakhov, Gregor Wawrzyniak, Neal E. Hartsough, Thulasi Gandhi, and Katsuyuki Taguchi "Characterization of a novel photon counting detector for clinical CT: count rate, energy resolution, and noise performance", Proc. SPIE 7258, Medical Imaging 2009: Physics of Medical Imaging, 725824 (12 March 2009); https://doi.org/10.1117/12.813915
PROCEEDINGS
9 PAGES


SHARE
Advertisement
Advertisement
RELATED CONTENT

Photon counting CT at elevated X ray tube currents ...
Proceedings of SPIE (March 18 2014)
Dual energy X ray photon counting using an LSO MPPC...
Proceedings of SPIE (August 26 2015)
Large-format photon-counting x-ray and gamma-ray imagers
Proceedings of SPIE (February 02 1993)
Cathode depth sensing in CZT detectors
Proceedings of SPIE (February 02 2004)
Optimizing CdTe detectors and ASIC readouts for high flux x...
Proceedings of SPIE (September 14 2011)

Back to Top