13 March 2009 Real time implementation of distortion corrections for a tiled EMCCD-based solid state x-ray image intensifier (SSXII)
Author Affiliations +
Proceedings Volume 7258, Medical Imaging 2009: Physics of Medical Imaging; 72583B (2009) https://doi.org/10.1117/12.813603
Event: SPIE Medical Imaging, 2009, Lake Buena Vista (Orlando Area), Florida, United States
Abstract
The new Solid State X-ray Image Intensifier (SSXII) is being designed based on a modular imaging array of Electron Multiplying Charge Couple Devices (EMCCD). Each of the detector modules consists of a CsI(Tl) phosphor coupled to a fiber-optic plate, a fiber-optic taper (FOT), and an EMCCD sensor with its electronics. During the optical coupling and alignment of the modules into an array form, small orientation misalignments, such as rotation and translation of the EMCCD sensors, are expected. In addition, barrel distortion will result from the FOTs. Correction algorithms have been developed by our group for all the above artifacts. However, it is critical for the system's performance to correct these artifacts in real-time (30 fps). To achieve this, we will use two-dimensional Look-Up-Tables (LUT) (each for x and y coordinates), which map the corrected pixel locations to the acquired-image pixel locations. To evaluate the feasibility of this approach, this process is simulated making use of parallel coding techniques to allow real-time distortion corrections for up to sixteen modules when a standard quad processor is used. The results of this simulation confirm that tiled field-of-views (FOV) comparable with those of flat panel detectors can be generated in ~17 ms (>30 fps). The increased FOV enabled through correction of tiled images, combined with the EMCCD characteristics of low noise, negligible lag and high sensitivity, should make possible the practical use of the SSXII with substantial advantages over conventional clinical systems. (Support: NIH Grants R01EB008425, R01NS43924, R01EB002873)
© (2009) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Christos Keleshis, K. R. Hoffmann, J. Lee, H. Hamwi, W. Wang, C. N. Ionita, D. R. Bednarek, A. Verevkin, and S. Rudin "Real time implementation of distortion corrections for a tiled EMCCD-based solid state x-ray image intensifier (SSXII)", Proc. SPIE 7258, Medical Imaging 2009: Physics of Medical Imaging, 72583B (13 March 2009); https://doi.org/10.1117/12.813603
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication and 1 patent.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Distortion

Sensors

Electron multiplying charge coupled devices

Computer simulations

X-rays

Image intensifiers

X-ray imaging

Back to Top