13 July 2009 Short and long term stability of the Diomed 630PDT laser evaluated with integrating sphere, power meter, and calorimeter
Author Affiliations +
Proceedings Volume 7380, Photodynamic Therapy: Back to the Future; 73804L (2009) https://doi.org/10.1117/12.823027
Event: 12th World Congress of the International Photodynamic Association, 2009, Seattle, Washington, United States
Abstract
The short and long term stability of the Diomed 630 PDT laser with attached fiberoptic microlens was evaluated by means of integrating sphere, power meter and a calorimetric system. The calorimeter system was designed as a thermal mug with absorbing media (dye and water). Both the tip of the irradiation fiber and the detection probe of a thermocouple thermometer were positioned inside the dye solution and stirred during the measurements. The calorimetric system yielded measurement results consistent with the other two methods, and similar long term variations were observed by all methods. With an indicated laser power of 1 W, the detectors' readings ranged from 0.66 to 1.29 W. For short term stability study, the deviation of laser output assessed by integrating sphere, power meter and calorimetric system were 0.3%, 0.1% and 2.8% with long term deviations of 13%, 7% and 9% respectively. This wide variation in the laser output implies the needs to establish quality control procedures involving measurements pre and post PDT procedures. The calorimetric system has been demonstrated to be a powerful tool for clinical laser QA and maintenance of the calibration factor of the detectors used in this work.
© (2009) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
C. Austerlitz, D. Campos, R. Allison, C. Sheng, C. Bonnerup, and C. Sibata "Short and long term stability of the Diomed 630PDT laser evaluated with integrating sphere, power meter, and calorimeter", Proc. SPIE 7380, Photodynamic Therapy: Back to the Future, 73804L (13 July 2009); https://doi.org/10.1117/12.823027
Lens.org Logo
CITATIONS
Cited by 2 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Integrating spheres

Power meters

Photodynamic therapy

Laser stabilization

Microlens

Fiber optics

Ocean optics

RELATED CONTENT

A quality assurance program for clinical PDT
Proceedings of SPIE (March 05 2018)
In vivo light dosimetry for pleural PDT
Proceedings of SPIE (February 18 2009)
Studies on the long term stability of fiber optic pressure...
Proceedings of SPIE (September 24 1996)

Back to Top