8 September 2009 Gain spectrum in gated x-ray MCPs
Author Affiliations +
Abstract
The gain spectrum in a gated multichannel intensifier output depends on the gain and spatial averaging. The spectrum affects the minimum signal that can be detected as well as the signal to noise in the detected images. We will present data on the gain-spectrum for the GXD detector, a gated x-ray detector to be used at the National Ignition Facility. The data was recorded on a cooled CCD detector, with an x-ray gating time of approximately 75 ps, selected from a range of 0.2 and 1 ns electrical pulse width determined by pulse forming modules were also used. The detector was characterized at the TRIDENT laser facility, using a 2.4 ns long x-ray at 4.75 keV. The x-rays were generated by the interaction of the focused Trident laser beam with a Titanium target.
© (2009) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
G. A. Kyrala, J. Oertel, T. Archuleta, Joe Holder, "Gain spectrum in gated x-ray MCPs", Proc. SPIE 7448, Advances in X-Ray/EUV Optics and Components IV, 74480S (8 September 2009); doi: 10.1117/12.825717; https://doi.org/10.1117/12.825717