You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
9 March 2010Detection of architectural distortion in prior mammograms using fractal analysis and angular spread of power
This paper presents methods for the detection of architectural distortion in mammograms of interval-cancer cases
taken prior to the diagnosis of breast cancer, using Gabor filters, phase portrait analysis, fractal dimension (FD),
and analysis of the angular spread of power in the Fourier spectrum. In the estimation of FD using the Fourier
power spectrum, only the distribution of power over radial frequency is considered; the information regarding
the angular spread of power is ignored. In this study, the angular spread of power in the Fourier spectrum is
used to generate features for the detection of spiculated patterns related to architectural distortion. Using Gabor
filters and phase portrait analysis, a total of 4224 regions of interest (ROIs) were automatically obtained from
106 prior mammograms of 56 interval-cancer cases, including 301 ROIs related to architectural distortion, and
from 52 mammograms of 13 normal cases. For each ROI, the FD and measures of the angular spread of power
were computed. Feature selection was performed using stepwise logistic regression. The best result achieved,
in terms of the area under the receiver operating characteristic curve, is 0.75 ± 0.02 with an artificial neural
network including radial basis functions. Analysis of the performance of the methods with free-response receiver
operating characteristics indicated a sensitivity of 0.82 at 7.7 false positives per image.