Paper
23 February 2010 Dynamic tracking of tendon elongation in ultrasound imaging
Mahta Karimpoor, Hazel Screen, Dylan Morrissey
Author Affiliations +
Abstract
The aim of this study was to evaluate the elongation of the Achilles tendon by looking at the changing position of Myo-Tendenious Junction (MTJ) using ultrasound during isometric contraction on an Isometric dynamometer. A sequence of ultrasound images in the form of movie, obtained from a unit operating at a frequency of 12MHz during isometric contraction, was analyzed offline using MATLAB to track the MTJ. This investigation has implemented important techniques for in vivo feature extraction of Achilles tendon. Prior to feature extraction, the images were filtered by anisotropic diffusion method and morphological enhancements. The cross correlation search algorithm with an adaptive mask was utilized to track MTJ by comparing adjacent segmented frames. The present method was studied on seventeen subjects, where it was able to measure the related movement accurately.
© (2010) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Mahta Karimpoor, Hazel Screen, and Dylan Morrissey "Dynamic tracking of tendon elongation in ultrasound imaging", Proc. SPIE 7625, Medical Imaging 2010: Visualization, Image-Guided Procedures, and Modeling, 76252P (23 February 2010); https://doi.org/10.1117/12.844623