21 July 2010 Monolithic Michelson Interferometer as ultra stable wavelength reference
Author Affiliations +
Abstract
Ultra-stable Monolithic Michelson interferometer can be an ideal reference for highprecision applications such as RV measurement in planet searching and orbit study. The advantages include wide wavelength range, simple sinusoidal spectral format, and high optical efficiency. In this paper, we report that a monolithic Michelson interferometers has been in-house developed with minimized thermal sensitivity with compensation tuning. With a scanning white light interferometer, the thermal sensitivity is measured ~ 6x10-7/°C at 550 nm and it decreases to zero near 1000 nm. We expect the wideband wavelength reference source to be stabilized better than 0.3 m/s for RV experiments
© (2010) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Xiaoke Wan, Xiaoke Wan, Jian Ge, Jian Ge, } "Monolithic Michelson Interferometer as ultra stable wavelength reference", Proc. SPIE 7734, Optical and Infrared Interferometry II, 77343N (21 July 2010); doi: 10.1117/12.857616; https://doi.org/10.1117/12.857616