27 August 2010 Detector dead time in particle counting detectors
Author Affiliations +
Abstract
Dead time in detectors limits the capability of the detector to work at high counting rates. A correction for dead time can be used to estimate the particle flux from a measurement. A motivation for operating detectors at high counting rates is to get the best statistical accuracy. It is therefore also important to correct measurements for the loss of statistical accuracy that occurs when dead time is present. When the source of particles has a time dependent variation in intensity, the conventional dead time correction may not apply. A new dead time correction that applies in this case is presented.
© (2010) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
David Laundy "Detector dead time in particle counting detectors", Proc. SPIE 7805, Hard X-Ray, Gamma-Ray, and Neutron Detector Physics XII, 780516 (27 August 2010); https://doi.org/10.1117/12.867128