Paper
21 February 2011 Attenuation of motion artifact in near infrared spectroscopy signals using a wavelet based method
Behnam Molavi, Guy D. Dumont, Babak Shadgan, Andrew J. Macnab
Author Affiliations +
Abstract
The sensitivity of Near Infrared Spectroscopy (NIRS) signals to motion artifact can limit practical applications of NIRS monitoring. We describe a wavelet based method for removing motion artifact from NIRS signals. This method was tested on monitoring data collected from leg muscle in 3 patients undergoing surgery to stabilize leg fractures. The results show an average of 18.36 dB attenuation in motion artifact energy in the NIRS signal for our test subjects when using the proposed method, without the introduction of significant distortion in the artifact-free regions of the signal.
© (2011) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Behnam Molavi, Guy D. Dumont, Babak Shadgan, and Andrew J. Macnab "Attenuation of motion artifact in near infrared spectroscopy signals using a wavelet based method", Proc. SPIE 7890, Advanced Biomedical and Clinical Diagnostic Systems IX, 78900M (21 February 2011); https://doi.org/10.1117/12.875741