21 February 2011 Automatic online spectral calibration of Fourier-domain OCT for robotic surgery
Author Affiliations +
Abstract
We present a new automatic spectral calibration (ASC) method for spectral Domain optical coherence tomography (SD-OCT). Our ASC method calibrates the spectral mapping of the spectrometer in SD-OCT, and does not require external calibrating light source or a commercial spectral analyzer. The ASC method simultaneously calibrates the physical pixel spacing of the A-scan in static and dynamic environments. Experimental results show that the proposed ASC method can provide satisfactory calibration for SD-OCT to achieve high axial resolution and high ranging accuracy, without increasing hardware complexity.
© (2011) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Xuan Liu, Xuan Liu, Marcin Balicki, Marcin Balicki, Russell H. Taylor, Russell H. Taylor, Jin U. Kang, Jin U. Kang, } "Automatic online spectral calibration of Fourier-domain OCT for robotic surgery", Proc. SPIE 7890, Advanced Biomedical and Clinical Diagnostic Systems IX, 78900X (21 February 2011); doi: 10.1117/12.875210; https://doi.org/10.1117/12.875210