Paper
17 February 2011 Radiofrequency circuit design and performance evaluation for small animal frequency-domain NIR fluorescence optical tomography
Chinmay Darne, Banghe Zhu, Yujie Lu, I-Chih Tan, John Rasmussen, Eva M. Sevick-Muraca
Author Affiliations +
Abstract
Herein we report on hardware development and evaluation for frequency-domain photon migration (FDPM) technique that is miniaturized for incorporation into a micro-CT gantry for hybrid CT/NIR/PET imaging. Immunity to endogenous optical properties and enhanced contrast associated with fluorophore lifetime is inherent to the FDPM measurements and enables unique opportunities for quantitative tomography when compared to the time independent (continuous wave) approach. A miniaturized radiofrequency (rf) circuitry has been developed in our laboratory for homodyne FDPM measurements that makes use of a single 100MHz oscillator to simultaneously launch optically modulated excitation light into a small animal as well as to modulate an NIR sensitive image intensifier for collection of fluorescent signals. The use of a single oscillator not only eliminates signal drift that otherwise results from the use of multiple oscillators individually driving both source and detector, but also reduces the circuit footprint for incorporation into the CT gantry. Herein, overall system performance parameters of signal-to-noise ratio, measurement precision, spatial resolution, modulation depth (ac/dc), excitation light rejection, and clinically relevant data acquisition times are presented for mouse phantom data. Image reconstruction of phantom data and integration of circuitry for hybrid CT/NIR/PET imaging is also presented towards the ultimate validation of NIR optical tomography using PET imaging as a "gold-standard" for quantification.
© (2011) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Chinmay Darne, Banghe Zhu, Yujie Lu, I-Chih Tan, John Rasmussen, and Eva M. Sevick-Muraca "Radiofrequency circuit design and performance evaluation for small animal frequency-domain NIR fluorescence optical tomography", Proc. SPIE 7896, Optical Tomography and Spectroscopy of Tissue IX, 789621 (17 February 2011); https://doi.org/10.1117/12.874946
Lens.org Logo
CITATIONS
Cited by 4 scholarly publications and 1 patent.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Near infrared

Charge-coupled devices

Luminescence

Optical tomography

Signal to noise ratio

Modulation

Oscillators

Back to Top