16 March 2011 New method to test the gantry, collimator, and table rotation angles of a linear accelerator used in radiation therapy
Stéphane Beaumont, Tarraf Torfeh, Romain Latreille, Yassine Ben Hdech, Jeanpierre Guedon
Author Affiliations +
Abstract
The precision of a medical LINear ACcelerator (LINAC) gantry rotation angle is crucial for the radiation therapy process, especially in stereotactic radio surgery, given the expected precision of the treatment and in Image Guided Radiation Therapy (IGRT) where the mechanical stability is disturbed due to the additional weight of the kV x-ray tube and detector. We present in this paper an extension of the Winston and Lutz test initially dedicated to control the size and the position of the isocenter of the LINAC and here adapted to test the gantry rotation angle with no additional portal images. This new method uses a test-object patented by QualiFormeD5 and is integrated in the QUALIMAGIQ software platform developed to automatically analyze images acquired for quality control of medical devices.
© (2011) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Stéphane Beaumont, Tarraf Torfeh, Romain Latreille, Yassine Ben Hdech, and Jeanpierre Guedon "New method to test the gantry, collimator, and table rotation angles of a linear accelerator used in radiation therapy", Proc. SPIE 7961, Medical Imaging 2011: Physics of Medical Imaging, 796153 (16 March 2011); https://doi.org/10.1117/12.878199
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Lithium

Collimators

Tungsten

Optical spheres

Image analysis

Medical imaging

Radiotherapy

Back to Top