You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
23 May 2011Picosecond laser pulses improves sensitivity in standoff explosive
detection
Portendo has in collaboration with FOI, the Swedish Defence Research Agency, developed a world-leading
technique of trace detection of explosives at standoff distance using Raman spectroscopy. The technology is
further developed in order to enhance the sensitivity of the method and to be able to extend the field of
applications. Raman scattering is a well-established technique able to detect substances down to single
micrograms at standoff distances, however, one of the obstacles limiting the detection possibilities is interfering
fluorescence, originating either from the substance itself or from the surrounding material. One main challenge
for this technology is thus to either omit the excitation of the fluorescent process altogether or to be able to
separate the two processes and only detect the Raman signal.
Due to the difference in the temporal behavior of the two processes - Raman scattering occurs in the order of
femtoseconds while fluorescence typically has a lifetime in the order of nanoseconds - one way to theoretically
separate them is to limit the measurement to as short time as possible, cutting off most of the emitted
fluorescence. The improvement depends on how much of the fluorescence can be omitted without decreasing
the Raman signal. Experimentally, we have verified this improvement in signal to noise ratio when using a laser
with picosecond pulses instead of nanosecond pulses, which has resulted in an improvement in SNR of up to 7
times for bulk ANFO. These results verify the predicted signal enhancement and suggest higher sensitivity for
standoff detection in future systems.
The alert did not successfully save. Please try again later.
Madeleine Åkeson, Markus Nordberg, Anneli Ehlerding, Lars-Erik Nilsson, Henric Östmark, Pierre Strömbeck, "Picosecond laser pulses improves sensitivity in standoff explosive detection," Proc. SPIE 8017, Detection and Sensing of Mines, Explosive Objects, and Obscured Targets XVI, 80171C (23 May 2011); https://doi.org/10.1117/12.883351