Paper
11 August 2011 Frequency calibration of terahertz time-domain spectrometers using absorption lines of carbon monoxide
Qing Sun, Yuqiang Deng, Jing Yu
Author Affiliations +
Abstract
We present a technique for calibrating spectrum measurement of terahertz time domain spectrometers (THz-TDS) using absorption lines of carbon monoxide (CO). The mean error between the peak positions measured by our THz-TDS and JPL database is 3.1 GHz. The results show that CO is suitable for frequency calibration of THz-TDS.
© (2011) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Qing Sun, Yuqiang Deng, and Jing Yu "Frequency calibration of terahertz time-domain spectrometers using absorption lines of carbon monoxide", Proc. SPIE 8195, International Symposium on Photoelectronic Detection and Imaging 2011: Terahertz Wave Technologies and Applications, 81950N (11 August 2011); https://doi.org/10.1117/12.900329