11 August 2011 The design of frequency-drift insensitive terahertz transmission measurement system
Author Affiliations +
Abstract
The precision of the conventional terahertz transmission measurement system with a parallel-plate sample will be reduced due to the interference effect, when the frequency of the THz source drifts. In order to solve such problem, a frequency-drift insensitive transmission measurement system composed of double-path is proposed in this paper, and an F-P interference model is built. The parameters are optimized, and the effects of the terahertz source's frequency drift on the precision of both the single- and double- path transmission measurement are analyzed and compared. The results demonstrated that such kind of measurement system could avoid the influence of the frequency-drift of the THz source, thus enhance the measurement accuracy.
© (2011) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Jianjun Liu, Jianjun Liu, Wuxi Ma, Wuxi Ma, Xiuhua Fu, Xiuhua Fu, Mengen Zhang, Mengen Zhang, Zhi Hong, Zhi Hong, } "The design of frequency-drift insensitive terahertz transmission measurement system", Proc. SPIE 8195, International Symposium on Photoelectronic Detection and Imaging 2011: Terahertz Wave Technologies and Applications, 81950X (11 August 2011); doi: 10.1117/12.900587; https://doi.org/10.1117/12.900587