9 February 2012 Miniature integrated optical coherence tomography (OCT) - ultrasound (US) probe for intravascular imaging
Author Affiliations +
Abstract
A miniature integrated optical coherence tomography (OCT) - ultrasound (US) probing system for real-time intravascular imaging has been developed. The outer diameter of the integrated probe is 0.69 mm, which is small enough for imaging in human coronary arteries. This probe, which has high resolution and deep tissue penetration, is designed to identify vulnerable atherosclerotic plaques in coronary arteries. The first in vivo images of a rabbit abdominal aorta obtained by the integrated OCT-US probe are presented.
© (2012) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Jiawen Li, Jiawen Li, Jiechen Yin, Jiechen Yin, Xiang Li, Xiang Li, Joe Jing, Joe Jing, David Mukai, David Mukai, Sari Mahon, Sari Mahon, Ahmad Edris, Ahmad Edris, Khiet Hoang, Khiet Hoang, K. Kirk Shung, K. Kirk Shung, Matthew Brenner, Matthew Brenner, Jagat Narula, Jagat Narula, Qifa Zhou, Qifa Zhou, Pranav Patel, Pranav Patel, Zhongping Chen, Zhongping Chen, } "Miniature integrated optical coherence tomography (OCT) - ultrasound (US) probe for intravascular imaging", Proc. SPIE 8207, Photonic Therapeutics and Diagnostics VIII, 82073X (9 February 2012); doi: 10.1117/12.905707; https://doi.org/10.1117/12.905707