30 January 2012 Picoliter-volume glucose concentration microsensor based on miniature abrupt-tapered Mach-Zehnder-interferometer
Zhi-Zheng Feng, Tsung-Hsun Yang, Nan-Kuang Chen, Yi-Ning Chen, Yu-Chung Chang
Author Affiliations +
Abstract
We report an ultracompact Mach-Zehnder index sensor with a device length of 438.6 μm using two micro-abrupt-tapers in a cladding-reduced strongly-guiding fiber based on a focused CO2 laser beam. The cladding of strongly-guiding fiber is chemically etched before irradiated by a focused CO2 laser beam. The index sensitivity is 600 nm/refractive index unit respectively at around 1.37 μm wavelength with a volume of 31.2 picoliter optical liquid trapping at one micro-abrupt-taper. A 72 picoliter glucose liquid with concentration of 200 mg/dL can lead to a red-shift of 0.8 nm at 1.3 μm.
© (2012) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Zhi-Zheng Feng, Tsung-Hsun Yang, Nan-Kuang Chen, Yi-Ning Chen, and Yu-Chung Chang "Picoliter-volume glucose concentration microsensor based on miniature abrupt-tapered Mach-Zehnder-interferometer", Proc. SPIE 8218, Optical Fibers and Sensors for Medical Diagnostics and Treatment Applications XII, 821807 (30 January 2012); https://doi.org/10.1117/12.904699