3 March 2012 The effect of breast positioning on breast compression in mammography: a pressure distribution perspective
Magnus Dustler, Ingvar Andersson, Daniel Förnvik, Anders Tingberg
Author Affiliations +
Abstract
The standard procedure at mammography is to compress the breast in order to improve image quality, better separate tissue components and reduce absorbed dose to the breast. Traditionally, compression guidelines have been based on applied force, rather than actual thickness reduction. Structures such as the pectoral muscle are stiffer than breast tissue and if compressed along with it, as in the MLO-projection, might absorb much of the applied force. This study investigated the difference in compression of breasts before and after they were repositioned to exclude 1 cm of the juxtathoracic part. Twenty-one women were included in the study. The distribution of compression pressure was measured using thin FSR (Force Sensing Resistor) pressure sensors attached to the compression paddle. Breast thickness and compression force were measured by the mammographic device. Compared to standard positioning the repositioned breasts were thinner by 4.4±2.3 mm (P < 0.001) (from 50.3 mm to 45.9 mm) and had a 12.3±24.5 cm2 (P = 0.032) larger area over which pressure was distributed (from 97.6 cm2 to 109.9 cm2), despite less of the breast being included in the projection. This indicates that the inclusion of the pectoral muscle and other juxtathoracic structures in the MLOprojection substantially affects pressure distribution and prevents proper compression of the breast. The results suggest that the exact positioning of the MLO-projection should be carefully evaluated in order to find a balance between breast compression and tissue inclusion.
© (2012) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Magnus Dustler, Ingvar Andersson, Daniel Förnvik, and Anders Tingberg "The effect of breast positioning on breast compression in mammography: a pressure distribution perspective", Proc. SPIE 8313, Medical Imaging 2012: Physics of Medical Imaging, 83134M (3 March 2012); https://doi.org/10.1117/12.905746
Lens.org Logo
CITATIONS
Cited by 11 scholarly publications and 3 patents.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Breast

Mammography

Sensors

Tissues

Image compression

Calibration

Image quality

Back to Top