You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
3 March 2012Development of a pinhole collimator SPECT-CT system using a CdTe detector sample
SPECT system with a co-registered x-ray CT image allows the fusion of functional information and morphologic
information. In this study, a pinhole collimator SPECT system sharing a CdTe detector with a CT was designed. Geant4
application for tomographic emission (GATE) v.6 was used for the design simulation. Sensitivity and spatial resolution
were measured to evaluate the system characteristics on an ideal condition. A pinhole collimator was designed to obtain
high spatial resolution and better sensitivity. The diameter of the collimator was 0.5 mm, and the distance from the center
of the field of view (CFOV) was set to 25 mm. The source to detector distance was set to 69 mm. The acquisition time of
each projection was 40 seconds, and 60 projections were obtained for tomographic image acquisition. The reconstruction
was performed using ordered subset expectation maximization (OS-EM) algorithms. The total system sensitivity was
measured in cps/kBq, and the spatial resolution was defined by the full width at half maximum (FWHM) from the point
spread function (PSF) of the point source image. Modulation transfer function (MTF) was also measured from the PSF.
The spatial resolution of the system calculated from the FWHM of Gaussian fitted PSF curve was 0.48 mm, and the
sensitivity of the system measured 0.354 cps/kBq from a Tc-99m point source. A phantom study was performed to verify
the dual modality system design. The system will be built as designed, and it can be applied for the pre-clinical imaging
system.