6 March 2012 NPS assessment of color medical displays using a monochromatic CCD camera
Author Affiliations +
Abstract
This paper presents an approach to Noise Power Spectrum (NPS) assessment of color medical displays without using an expensive imaging colorimeter. The R, G and B color uniform patterns were shown on the display under study and the images were taken using a high resolution monochromatic camera. A colorimeter was used to calibrate the camera images. Synthetic intensity images were formed by the weighted sum of the R, G, B and the dark screen images. Finally the NPS analysis was conducted on the synthetic images. The proposed method replaces an expensive imaging colorimeter for NPS evaluation, which also suggests a potential solution for routine color medical display QA/QC in the clinical area, especially when imaging of display devices is desired.
© (2012) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Hans Roehrig, Hans Roehrig, Xiliang Gu, Xiliang Gu, Jiahua Fan, Jiahua Fan, } "NPS assessment of color medical displays using a monochromatic CCD camera", Proc. SPIE 8318, Medical Imaging 2012: Image Perception, Observer Performance, and Technology Assessment, 83181Q (6 March 2012); doi: 10.1117/12.913749; https://doi.org/10.1117/12.913749
PROCEEDINGS
7 PAGES


SHARE
RELATED CONTENT


Back to Top