Paper
17 October 2012 Multiplexed regenerated Fiber Bragg Gratings for high temperature measurement
Author Affiliations +
Proceedings Volume 8421, OFS2012 22nd International Conference on Optical Fiber Sensors; 842123 (2012) https://doi.org/10.1117/12.970624
Event: OFS2012 22nd International Conference on Optical Fiber Sensor, 2012, Beijing, China
Abstract
High temperature gradient of a conventional tubular furnace is characterized using a single fiber sensing line with wavelength-multiplexed short-length regenerated Fiber Bragg Gratings. The multiplexed gratings are simultaneously regenerated using a high temperature annealing process. Temperature calibration from ambient temperature up to 850°C is conducted leading to a standard deviation of 0.33°C after polynomial fitting of the wavelength shift with temperature.
© (2012) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
G. Laffont, R. Cotillard, and P. Ferdinand "Multiplexed regenerated Fiber Bragg Gratings for high temperature measurement", Proc. SPIE 8421, OFS2012 22nd International Conference on Optical Fiber Sensors, 842123 (17 October 2012); https://doi.org/10.1117/12.970624
Lens.org Logo
CITATIONS
Cited by 4 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Fiber Bragg gratings

Multiplexing

Temperature metrology

Calibration

Reflectivity

Transducers

Annealing

Back to Top